新填充的色譜柱不能馬上使用還需要進行老化處理。老化的目的有兩個,一是為了除去填充物中的殘余溶劑,和某些揮發(fā)性雜質(zhì),另一個目的是促進固定液均勻的、牢固分布在單體的表面上。
新填充的色譜柱老化的方法:----把柱子與汽化室連接,與檢測器一端要斷開,以氮氣為載氣,流速是正常的一半即可,溫度選擇固定液的使用溫度,老化時間大約20小時,老化完成后將儀器溫度降至近室溫關(guān)閉色譜儀,待儀器溫度恢復室溫再將色譜柱連接到檢測器上(老化時接汽化室的一端接在檢測器上),開機,在使用溫度下看基線是否平穩(wěn),如果平穩(wěn)色譜柱就算老化好了,否則要繼續(xù)老化。
英文翻譯:
Newly filled columns cannot be used immediately and need to be aged.There are two purposes of aging ,one is to completely remove the residual solvent and some volatile impurities in the filler, and the other is to promote the uniform and firm distribution of the fixing solution on the surface of the monomer.
New filling method of chromatographic column aging: a pair of which is connected with vaporzing chamber, and the detector at one end to disconnect , air as carrier gas ,the velocity is half the normal can , choose fixed liquid highest use temperature, aging time of about 20 hours after aging turn off the instrument temperature dropped to near room temperature and chromatograph instrument temperature recovery at room temperature and then connect the chromatographic column on the detector(when aging, one end of the vaporization chamber is best
connected to the detector) ,start up , see wheter the baseline is stable at the use temperature, if the stable column is aging, otherwise continue to aging.