氫氣發(fā)生器的安裝與使用
今日新聞 日期:2022/5/17
1、將儀器從包裝箱內(nèi)取出,檢查有無因運輸不當(dāng)而損壞,核對儀器備件,合格證及保修卡是否齊全。
2、連接儀器氣路管,保證接頭不漏氣。
3、打開儀器的水桶蓋,加入純水、去離子水,不要超過觀察水位線,也不要低于下限水位線,擰緊上蓋。
4、接通電源,打開電源開關(guān),電源開關(guān)紅燈亮,產(chǎn)氣燈綠燈亮。輸出流量液晶顯示屏顯示,顯示數(shù)為,400,壓力表指針由0Mpa上升到0.4Mpa儀器可正常使用。
5、儀器自帶有干燥管一套。使用過程中注意觀察干燥管中的硅膠是否變色,如變色部分超過整部分的一半時應(yīng)及時更換變色硅膠。更換時應(yīng)在無壓力情況下進行,打開干燥管頂部上蓋,取出內(nèi)管把變色部分倒掉加入新的變色硅膠或再生后的硅膠。然后把內(nèi)管放入干燥管中確定放好,擰緊上蓋,確保不漏氣。
6、儀器使用一段時間后,水桶內(nèi)的純水會變少,變混濁應(yīng)及時加入或更換純水。建議:3-6個月水桶內(nèi)的水全部放掉,清洗水桶后加入新水。儀器運輸時將水桶內(nèi)的水放掉,放水方法:將儀器背后的放水硅膠管拉出,拿掉尼龍堵即可。
英文翻譯:
Installation and use of hydrogen generator
News of the day Date:2022/5/17
1.Take the instrument out of the packing box,check whether it is damaged due to improper transportation,and check whether the spare parts,certificates and warranty cards of the instrument are complete.
2.Connect the gas pipe of the instrument to ensure that the joint is not lesking.
3.open the bucket cover of the instrument,add pure water and deionized water,do not exceed the observation water level,nor lower than the lower limit water level,press the upper cover.
4.Turn on the power supply and turn on the power switch, The power switch is red and the gas production light is green. LCDdisplay of output flow,display number is,400,pressure gauge pointer rises from OMpa to0.4mpa instrument can be used normally.
5.the instrument comes with a set of drying pipe.In the process of use,pay attention to observe whether the silica gelin the drying tube is discolored. if the discolored part is more than half of the whole part,the discolored silica gel should be replaced in time.Replacement should be carride out under no pressure,open the top cover of the drying tube,take out the inner tube ti pour the discoloration part and add new discoloration silica gel or recycled silica gel.Then put the inner tube into the drying tube to ensure that it is properly placed,and expand the upper cover to ensure that there is no air leakage.
6.After the instrument is used for a period of time,the pure water in the bucket will become less,and the pure water should be added or replaced in time.Suggestion:Drain all the water in the bucket after 3-6 months,and add new water after cleaning the bucket. During the transportation of the instrument,drain the water in the bucket.Method of water discharge:pull out the silicone tube of water discharge at the back of the instrument and remove the nylon plug.